Objective: We aimed to establish high performance liquid chromatography (HPLC) method to measure plasma melatonin levels easily, sensitively and reliably.
Methods: We used fluorescence detection, 75 mM sodium acetate/acetonitrile (72/28, v/v)(pH 5) as a mobile phase, isocratic method, 275 and 345 nm excitation and emission wavelengths, respectively. Samples were extracted using solid phase extraction column (Waters Oasis HLB 1 cc cartridge).
Results: The recovery was found as 97.2%. Inter-assay and intra-assay coefficients of variation (CV) were 4.02% and 5.41%, respectively. Detection limit was 0.25 pg/mL and linear range was up to 1000 pg/mL.
Conclusion: These findings suggest that our method is suitable to measure plasma melatonin levels.

Key words: HPLC; Melatonin; SPE



Share this Article


Advertisement
American Journal of Physiology, Biochemistry and Pharmacology

SUBMIT YOUR ARTICLE NOW


ScopeMed.com
ScopeMed Home
Follow ScopeMed on Twitter
BiblioCAM
Author Tools
eJPort Journal Hosting
About ScopeMed
License Information
Terms & Conditions
Privacy Policy
Contact Us

The articles in Scopemed are open access articles licensed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License (https://creativecommons.org/licenses/by-nc-sa/4.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.
ScopeMed is a Database Service for Scientific Publications. Copyright ScopeMed Information Services.
Scopemed Buttons